NEW Systec High-Flow Degasser, 1000 mL/min - 0001-6706

  • Model: 0001-6706
  • Manufactured by: Systec


Call for Price

SYSTEC® High Flow Degasser


SYSTEC’s High Flow Degasser module includes a high efficiency, non-porous membrane and controlled vacuum pump for liquid vacuum degassing applications.

Features 
  • Removes dissolved gasses and bubbles from fluidic lines to improve dispense precision
  • Longest lifetime pump and membrane materials are selected to reduce maintenance and provide best TCO
  • Compatible with broad range of fluids including aqueous salts, organics, detergents and sterilizing agents (pH range: 2 to 13)
  • Highest efficiency membrane reduces unit size to increase lab space
  • Tested and calibrated assembly is ready for installation

TECHNICAL SPECIFICATIONS

Operational flow rate (mL/min): 0 - 1,000

Pressure Drop (@ 20 °C & Max Flow): < 15 kPa

Internal Volume: < < 45 mL

Degassing Efficiency (Removed O2% in DI Water @ Max Flow): > 50%

Number of Channels: 1

Wetted Materials: Polysulfone, silicone rubber, polypropylene, Tygon S-50-HL, medical grade adhesives

External Fluidic Ports: 3/16 inch (4.8 mm) ID Barb Fittings

Maximum Operating Pressure: < 2 Barg

Maximum Temperature: 40 °C


PUMP CONTROL SPECIFICATIONS

Power Requirements: 24 VDC @ 0.75 Amp max. (<5 Watt typical)

Vacuum Sensor Calibration Accuracy: (NIST Traceable) ±5 mmHg absolute over an ambient operating temperature range of 10-35°C.

Closed-Loop Control Setpoint: 120 mmHg (16kPa) absolute (pump runs at high RPM until
near setpoint, then speed is varied to maintain a value of 120 mmHg (16kPa) absolute – load independent).

LED Indicators
Single Bi-Color
Power on, vacuum above control range: Amber– Blinking
Vacuum reaches upper control range: Green-Blinking
Vacuum reaches control set point: Green—Solid
Error Condition, Shutdown: Amber—Solid

Errors Detected
Pumpdown: Unable to reach 120 mmHg (16kPa) in 10 minutes.
High Vacuum: Vacuum > 200 mmHg (26.7kPa) for > 10 min.
Sensor Signal: Sensor >800 mmHg (106.7kPa) or <10 mmHg (1.3kPa).


VACUUM PUMP TECHNICAL DATA

Air Flow (no vacuum): 580 SCCM @ 250 RPM; 140 SCCM @ 60 RPM

Vacuum Performance Specification: <150 mmHg (20kPa) @ 120 RPM & 1 SCCM air flow (Test Criteria)

Pump Head Continuous Purge Air Flow Rate: 11 SCCM

Materials in Fluid Path: Stainless Steel, Polypropylene, PTFE, EPDM Rubber, PPS

Expected Lifetime: >6 years (continuous run @ 100 RPM 12 hours/day 365 days/year)


International Power Adapter Included
AC adapter: 100-240V AC, 1A, 50-60Hz
(includes adapters for North America, Japan, UK, Continental Europe, Australia)